Magnetic induction tomography: hardware for multi-frequency measurements in biological tissues

Hermann Scharfetter, Helmut Lackner, J. Rosell

Physiological Measurement · 2001 · 131 citations · 12 references

Concepts

TL;DR

Magnetic induction tomography is a contactless technique that maps tissue conductivity by detecting voltage changes induced in a receiver coil when an alternating magnetic field is perturbed by conductive tissue, with physiologically relevant frequencies requiring a signal‑to‑carrier ratio as low as 10⁻⁷ and potential for further resolution improvement. The study introduces a new MIT hardware comprising a coil system with planar gradiometers and a high‑resolution phase detector. The hardware uses planar gradiometers and a phase detector to operate over 20–370 kHz, extendable to 1 MHz. The system achieves an SCR of 2.5×10⁻⁵ (20 dB SNR at 150 kHz, 100 ms acquisition) and experimentally demonstrates feasible online spectroscopy of tissue conductivity in the β‑dispersion range, opening possibilities for non‑invasive brain oedema monitoring.

Abstract

Magnetic induction tomography (MIT) is a contactless method for mapping the electrical conductivity of tissue. MIT is based on the perturbation of an alternating magnetic field by a conducting object. The perturbation is detected by a voltage change in a receiver coil. At physiologically interesting frequencies (10 kHz-10 MHz) and conductivities (<2 S m-1) the lower limit for the relative voltage change (signal/carrier ratio = SCR) to be resolved is 10-7-10-10. A new MIT hardware has been developed consisting of a coil system with planar gradiometers and a high-resolution phase detector (PD). The gradiometer together with the PD resolves an SCR of 2.5×10-5 (SNR = 20 dB at 150 kHz, acquisition speed: 100 ms). The system operates between 20 and 370 kHz with the possibility of extending the range up to 1 MHz. The feasibility of measuring conductivity spectra in the β-dispersion range of biological tissues is experimentally demonstrated. An improvement of the resolution towards SCR = 10-7 with an SNR of ⩾20 dB at frequencies >100 kHz is possible. On-line spectroscopy of tissue conductivity with low spatial resolution appears feasible, thus enabling applications such as non-invasive monitoring of brain oedema.

References

12