Journal of High Resolution Chromatography · 1992 · 36 citations · 8 references
EngineeringPharmaceutical AnalysisBiomemsMicrofabricationTransport PhenomenaCapillary PressureColumn TechnologyCapillary ColumnsCapillary Liquid ChromatographyColumn Packing TechnologyGas Pressure EffectsPacked ColumnsPacking DensityPacking Efficiency
Abstract A dry‐packing method has been developed which enables the preparation of packed capillary columns for micro HPLC from 250 μm i.d. tubing and 5 μm packing materials. Pressurized gases, such as hydrogen and argon, were used to transport the packing media, and ethanol or methanol were used as discharge agents. By changing the gas pressure, the packing density could be easily adjusted. It was found that, within experimental limits, the higher the packing density, the greater the column efficiency. Comparison between dry‐ and slurrypacked columns showed that the former had greatly improved stability; the efficiency of dry‐packed columns was about the same as, or even better than, that obtained by slurry‐packing.
8
A study of micro-high-performance liquid chromatography
Daisuke Ishii, Kazuki Asai, K. Hibi et al. · Journal of Chromatography A · 1977 · 323 citations