Metrologia · 2011 · 57 citations · 38 references
Natural SciencesPhysicsAtomic PhysicsInstrumentationNuclear InstrumentationAtomic ClocksAtomic SpectroscopyCold Atom PhysicsLaser TrappingAtomic Frequency StandardsFrequency AccuracyFrequency StabilityMagneto-optical EffectsMagneto-optical PropertiesMagneto-optical TrapsRubidium Frequency StandardsSystematic Frequency Shifts
The design, operating parameters and the accuracy evaluation of the NPL Rb atomic fountain are described. The atomic fountain employs a double magneto-optical arrangement that allows a large number of 87Rb atoms to be trapped, a water-cooled temperature-stabilized interrogation region and a high quality factor interrogation cavity. From the uncertainties of measured and calculated systematic frequency shifts, the fractional frequency accuracy is estimated to be 3.7 × 10−16. The fractional frequency stability, limited predominantly by noise in the local oscillator, is measured to be 7 × 10−16 after one day of averaging. Based on the proposed quasi-continuous regime of operation of the fountain, the accuracy of the Rb standard of 5 × 10−17 reachable in two days of averaging is predicted.
38
A Molecular Beam Resonance Method with Separated Oscillating Fields
Norman F. Ramsey · Physical Review · 1950 · 1K citations
Natural Sciences, Chemistry, Physics +15
G. Breit, I. I. Rabi · Physical Review · 1931 · 658 citations
Quantum Projection Noise in an Atomic Fountain: A High Stability Cesium Frequency Standard
G. Santarelli, Ph. Laurent, P. Lemonde et al. · Physical Review Letters · 1999 · 557 citations
Natural Sciences, Physics, Photonics +14
R. Wynands, S. Weyers · Metrologia · 2005 · 365 citations