Magnetic Resonance in Medicine · 2009 · 98 citations · 29 references
MedicineBiomedical EngineeringRadiologyBiomedical ImagingDiagnostic ImagingSpectroscopyBrain ImagingMagnetic Resonance ImagingMagnetic Resonance Signal FormationMagnetic Resonance SpectroscopyMedical ImagingNeuroimaging PhysicsAdiabatic PulsesBrain RegionsChemical Shift ImagingMagnetic Field DistributionMagnetic Resonance Spectroscopic ImagingPower DepositionTransceiver ArchitecturesTransceiver Design
Spectroscopic imaging of the human head at short echo times (<or=15 ms) typically requires suppression of signals from extracerebral tissues. However, at 7 T, decreasing efficiency in B1+ generation (hertz/watt) and increasing spectral bandwidth result in dramatic increases in power deposition and increased chemical shift registration artifacts for conventional gradient-based in-plane localization. In this work, we describe a novel method using radiofrequency shimming and an eight-element transceiver array to generate a B1+ field distribution that excites a ring about the periphery of the head and leaves central brain regions largely unaffected. We have used this novel B1+ distribution to provide in-plane outer volume suppression (>98% suppression of extracerebral lipids) without the use of gradients. This novel B1+ distribution is used in conjunction with a double inversion recovery method to provide suppression of extracerebral resonances with T1s greater than 400 ms, while having negligible effect on metabolite ratios of cerebral resonances with T1s>1000 ms. Despite the use of two adiabatic pulses, the high efficiency of the ring distribution allows radiofrequency power deposition to be limited to 3-4 W for a pulse repetition time of 1.5 sec. The short echo time enabled the acquisition of images of the human brain, displaying glutamate, glutamine, macromolecules, and other major cerebral metabolites.
29
Ulrich Katscher, Peter Börnert, Christoph Leussler et al. · Magnetic Resonance in Medicine · 2002 · 703 citations · Full text