Applied Spectroscopy · 1985 · 57 citations · 29 references
Natural SciencesChemical EngineeringChemistryAnalytical ChemistryElemental AnalysisSpectroscopyAbsorption SpectroscopyAtomic Absorption SpectroscopyAtomic SpectroscopyElemental Composition AnalysisNebulizer SystemsAnalytical SensitivityAnalytical Sensitivity AssessmentContinuous FlowFlame Atomic Absorption SpectrometryFlame Atomic Absorption SpectroscopyHydride Generation Atomic Absorption SpectrometryHydride Generation Atomic SpectroscopySensitivity
The design and performance characteristics of a continuous-flow vapor generation accessory for atomic absorption spectrophotometry are reported. When used with a conventional atomic absorption spectrophotometer, the accessory provides analytical sensitivity for mercury, arsenic, selenium, antimony, bismuth, tellurium, and tin equal or superior to that of a batch-type vapor generation device. Sample presentation, sample consumption, precision of measurement, and simplicity of operation are similar to those of flame atomic absorption with a pneumatic nebulizer. Details of the construction of the continuous-flow manifold, gas/liquid separator, and absorption cells are discussed, and the effects of reagent concentrations and carrier gas flow rates are shown. Characteristic concentrations and analytical curves for the above elements are also presented.
29
Michael Thompson, B. Pahlavanpour, Sarah‐Jane Walton et al. · The Analyst · 1978 · 258 citations
Preservation of dilute mercury solutions
Cyrus Feldman · Analytical Chemistry · 1974 · 190 citations