Reduction in ischemic vascular complications with a hydrophilic-coated intra-aortic balloon catheter

Kenneth J. Winters, Stacy C. Smith, Marc Cohen, Christine Kopistansky, Ruth McBride

Catheterization and Cardiovascular Interventions · 1999 · 12 citations · 13 references

Concepts

Abstract

Hydrophilic coatings improve the handling characteristics of angioplasty equipment, but until recently this technology has not been available for intra-aortic balloon (IAB) catheters. To determine whether a new hydrophilic coating is associated with a reduction in IAB-related complications, we prospectively evaluated 188 patients undergoing insertion of this new IAB catheter. Complications related to IAB insertion were recorded and compared to data from 682 patients previously treated with an uncoated IAB catheter of the same shaft diameter and balloon size. By multivariate analysis the hydrophilic-coated IAB catheter was associated with a 72% reduction in ischemic vascular complications (relative risk 0.28, 95% CI 0.08-0.96, P= 0.04). There were also trends toward reductions in IAB rupture and the need for vascular surgical repair for bleeding or ischemia. While further investigation is warranted, hydrophilic coatings that reduce the coefficient of friction during IAB catheter insertion may also reduce subsequent ischemic vascular complications.

References

13