Review of Scientific Instruments · 1999 · 10 citations · 10 references
Natural SciencesPhysicsPlasma PhysicsRadiation DetectionRadiation MeasurementRadiographic ImagingX-ray Data AnalysisX-ray DetectionX-ray OpticsGainHigh-resolution X-ray ImagingQuantitative Imaging CalibrationSoft X-ray OpticsTime ResolutionX-ray DiagnosticsX-ray Imaging PhysicsZ-pinch Plasmas
A multiframe, gated pinhole system capable of quantitative acquisition for pulsed soft x rays is described and tested. The system based on a gated microchannel plate (MCP) is employed to observe the evolution of z-pinch plasma in a plasma focus facility with a time resolution of ∼220 ps. The quantitative relationship between x-ray source intensity and the recorded images has been investigated. To make a quantitative measurement, the phosphor screen current was measured to calculate the total electrons output from the MCP, which is proportional to both the incident x-ray intensity and the intensity of the recorded images. Furthermore, by taking into account the pinhole geometry, MCP gain and system spectral response, a quantitative calibration of the x-ray images has been established. We have employed the system to observe the plasma evolution in a plasma focus facility. An order of ∼1016 photons/(s mm2 mrad2) soft x-ray emission within 6–14 Å was observed in a neon puffed experiment with time resolved four successive frames.
10