Concepedia

Publication | Open Access

Evaluation of the Diagnostic Accuracy of CareStart G6PD Deficiency Rapid Diagnostic Test (RDT) in a Malaria Endemic Area in Ghana, Africa

38

Citations

27

References

2015

Year

Abstract

The outcome of this study suggests that the diagnostic performance of the CareStart G6PD deficiency RDT kit was high and it is acceptable at determining the G6PD deficiency status in a high malaria endemic area in Ghana. The RDT kit presents as an attractive tool for point-of-care G6PD deficiency for rapid testing in areas with high temperatures and less expertise. The CareStart G6PD deficiency RDT kit could be used to screen malaria patients before administration of the fixed dose primaquine with artemisinin-based combination therapy.