Review of Scientific Instruments · 1987 · 50 citations · 9 references
Natural SciencesElectrical EngineeringPhysicsInstrumentationNuclear InstrumentationPlasma PhysicsCharged Particle DetectionHeavy-ion PhysicsPlasma DiagnosticsPlasma ElectrodynamicsPlasma OscillationsBeam InstrumentationCalibration ProceduresIon Beam DiagnosticsPlasma Potential Measurements
The operation of a Proca and Green type 30° parallel plate electrostatic energy analyzer is modeled in a new manner that permits high-resolution heavy ion beam probe measurements of fluctuating plasma potential. Systematic calibration procedures permit detection of potential changes smaller than 0.01% of the probing beam energy at frequencies up to a megahertz. Most recent applications of beam probes have made use of this new capability.
9
A direct measurement of plasma space potential
F. C. Jobes, R. L. Hickok · Nuclear Fusion · 1970 · 128 citations
Space Potential Distribution in the ISX-B Tokamak
G. A. Hallock, Joseph Mathew, William C. Jennings et al. · Physical Review Letters · 1986 · 81 citations
TEXT heavy ion beam probe system
P. M. Schoch, John Forster, William C. Jennings et al. · Review of Scientific Instruments · 1986 · 48 citations
Elmo Bumpy Torus heavy ion probe
F.M. Bieniosek, P.L. Colestock, K. A. Connor et al. · Review of Scientific Instruments · 1980 · 40 citations