Biomedical Optics Express · 2012 · 27 citations · 18 references
EngineeringBioengineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingFluorescence ImagingFluorescence SpectroscopyIntravascular ImagingMedical Imaging InstrumentationOptical DiagnosticsOptical SpectroscopyFluorescence Emission SpectraFluorescence LifetimeFluorescence Lifetime ImagingFluorescence Lifetime Imaging Of TissueFluorescence Lifetime MeasurementsSpectral Fluorescence ImagingTime-resolved Fluorescence ImagingTime-resolved Fluorescence Spectroscopy
This study describes a scanning time-resolved fluorescence spectroscopy (TRFS) system designed to continuously acquire fluorescence emission and to reconstruct fluorescence lifetime images (FLIM) from a luminal surface by using a catheter-based optical probe with rotary joint and pull-back device. The ability of the system to temporally and spectrally resolve the fluorescence emission from tissue was validated using standard dyes and tissue phantoms (e.g., ex vivo pig aorta phantom). Current results demonstrate that this system is capable to reliably resolve the fluorescence emission of multiple fluorophores located in the lumen; and suggest its potential for intravascular detection of distinct biochemical features of atherosclerotic plaques.
18
Fluorescence Lifetime Standards for Time and Frequency Domain Fluorescence Spectroscopy
Noël Boens, Wenwu Qin, Nikola Basarić et al. · Analytical Chemistry · 2007 · 472 citations · Full text