Review of Scientific Instruments · 2001 · 67 citations · 22 references
EngineeringBiomedical EngineeringNuclear EngineeringBiomedical ImagingDiagnostic ImagingMicrofabricationMagnetic CharacterizationMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMagnetic Resonance Signal FormationMagnetic Resonance SpectroscopyCoiled CoilsNuclear Magnetic Resonance ParametersNuclear Magnetic Resonance SpectroscopyNuclear Magnetic Resonance StudiesRadiofrequency Coil DesignRadiofrequency CoilsReceiver Coil MiniaturizationTurn Spacing
We report the design and testing of a nuclear magnetic resonance (NMR) microcoil receiver apparatus, employing solenoidal microreceiver coils of dimensions of tens to hundreds of microns, using applied field of 9 T (proton resonance frequency 383 MHz). For the smallest receiver coils we attain sensitivity sufficient to observe proton NMR with signal to noise (S/N) one in a single scan applied to a ∼10 μm3 (10 fl) water sample, containing 7×1011 total proton spins. We also test the dependence of the S/N on important coil parameters, including coil composition and resistivity, turn spacing, and lead lengths.
22
Principles of Nuclear Magnetic Resonance in One and Two Dimensions
Richard R. Ernst, Geoffrey Bodenhausen, Alexander Wokaun et al. · Physics Today · 1989 · 4.2K citations
Natural Sciences, Chemistry, Physics +17
High-Resolution Microcoil1 H-NMR for Mass-Limited, Nanoliter-Volume Samples
Dean L. Olson, Timothy L. Peck, Andrew Webb et al. · Science · 1995 · 498 citations