Aesthetic Surgery Journal · 1999 · 13 citations · 10 references
MedicineSurgeryImplant SurgeryReconstructive SurgerySurgical InstrumentationBreast AugmentationBreast SurgeryBreast Volume EffectsFill Volume EffectsImplant-based Breast AugmentationSaline Breast ImplantsSilicone Gel Breast Implants
Background: The etiology for reported variations in saline solution-filled breast implant deflation rates (1% to 34%) remains obscure. Objective: The study's purpose was to determine whether the volume used in saline solution-filled breast implants affects the rate of spontaneous deflation. Methods: We performed a retrospective analysis of 345 saline solution-filled implants made by a single manufacturer and used for primary augmentations during a 7-year period. Results: We found a spontaneous deflation rate of 7.25%. The deflation rate was determined to be inversely proportional to the implant fill volume. Conclusions: Our study suggests that inflation of implants to the upper limits of the manufacturer's recommendations, or even slightly exceeding that amount, may reduce the rate of spontaneous deflation.
10
?No-Touch? submuscular saline breast augmentation technique
R MLADICK · Aesthetic Plastic Surgery · 1993 · 147 citations
Medicine, Plastic Surgery, Surgery +11
Saline inflatable prostheses: 14 years' experience
David M. Lavine · Aesthetic Plastic Surgery · 1993 · 63 citations
Medicine, Surgery, Prosthetics +2