Decomposition of Copper Compounds in X-Ray Photoelectron Spectrometers
162
Citations
8
References
1984
Year
Materials ScienceInorganic ChemistryCopper CompoundsX-ray GunX-ray SpectroscopyEngineeringHealth SciencesSpectroscopyX-ray DiffractionX-ray TechnologySix Cu CompoundsAnalytical ChemistryChemistryInstrumentationElemental MetalElemental CharacterizationWater-cooled X-ray HousingX-ray Imaging
Six Cu compounds were analyzed in three different x-ray photoelectron spectrometers (AEI-ES200, LHS-10, and HP-5950A), each equipped with a different type of AI Kα x-ray source, to determine the cause of reduction of copper salts in photoelectron spectrometers The monochromatic source used in the HP 5950A spectrometer did not cause sample degradation Significant sample degradation was observed in the AEIES200, which was attributed to the radiative heating from the x-ray gun The LHS-10 spectrometer, which utilizes a water-cooled x-ray housing, afforded significantly less sample reduction than the AEI ES200 spectrometer