Japanese Journal of Applied Physics · 2013 · 40 citations · 21 references
Natural SciencesBiomedical EngineeringAcousticsUltrasonographyUltrasoundAcoustic ImagingDiagnostic UltrasonographyTransducer TechnologyUltrasonicsUltrasound ImagingUltrasound PhysicsAmplitude CorrectionHeterogeneous Media EffectsHigh-intensity Focused UltrasoundNumerical SimulationPhase CorrectionPhased Array TransducersTransverse Wave PropagationUltrasonic Imaging
A focus control method based on time reversal aided by numerical simulation was implemented to correct focal errors in heterogeneous media by using a 56-element phased array transducer at a frequency of 2 MHz. Two types of acrylic phantom were employed to mimic the acoustic heterogeneities in the human body. Focal errors were significantly reduced by applying the phase correction. Moreover, the effects of amplitude correction, array configuration of the transducer, and transverse wave propagation in an elastic body were examined to improve the focal quality. The results suggested that the ultrasound was effectively converged on the target by correcting the amplitude of the ultrasound, and the accuracy of the focus control was enhanced by changing the array configuration and taking into account the shear elasticity of the elastic body in the simulation.
21
Effect of high-intensity focused ultrasound on human prostate cancer in vivo.
Stephan Madersbacher, M. Pedevilla, L Vingers et al. · PubMed · 1995 · 370 citations