Geostandards and Geoanalytical Research · 1980 · 10 citations · 16 references
Natural SciencesChemistryGeologyAnalytical ChemistryElemental AnalysisSpectroscopyEnvironmental Elemental AnalysisHeavy Metal AnalysisHeavy Metal ContaminationTrace Element ChemistryTrace Metal ChemistryDetection Limit AssessmentHydride FormationHydride Generation Atomic Absorption SpectrometryInstrument Detection LimitMethod Detection LimitUltratrace Element AnalysisUltratrace Metal AnalysisUltratrace Metal Determination
A method is described for the determination of mercury at nanogram levels in geological materials by flameless atomic absorption spectroscopy after metal vapor generation. Mercury is converted to a soluble form by H 2 SO 4 ‐HNO 3 attack in a teflon bomb at 180°C for 90 minutes. Metal Hg vapor is obtained by reaction with combined NaBH 4 ‐NaOH using the Perkin‐Elmer MHS ‐ 10 hydride system and determined at room temperature. Mercury can be easily determined in most silicate materials. The sensitivity of the method depends essentially on the purity of the reagents and contaminations during handling rather than on instrumental capabilities. A detection limit of about 10 ng/g Hg for 0.5 g samples has been achieved with the use of normal “suprapure” reagents. Mercury contents on some USGS, CRPG and ANRT reference samples are reported.
16