Concepedia

Publication | Open Access

Absolute Measurement of Loss Angle Using a Toroidal Cross Capacitor

25

Citations

8

References

1978

Year

Abstract

An absolute measurement of loss angle has recently been completed at NBS. The measurement utilized a special toroidal cross capacitor in which the effects of dielectric films are greatly attenuated. The resulting unit of loss angle has an estimated uncertainty of 0.02 μrad at an optimum frequency of 1592 Hz.