EngineeringBiomedical EngineeringBiomedical ImagingBiomedical OpticsDiagnostic ImagingOptical Coherence TomographyOptical TomographyOral Cavity CancerOral Disease DiagnosticsAbnormal Tissue DetectionAxial ResolutionEx Vivo ImagingEx Vivo TomographyExponential Curve FittingOral Lesion DetectionOral Tumor ImagingSensitivity
We demonstrate the ex vivo imaging of an oral cancerous sample with a swept-source optical coherence tomography (SS-OCT) system. With the axial resolution of 8 microm in free space and system sensitivity of 108 dB, we can well distinguish the normal and abnormal tissue portions in a sample. In particular, we analyze the lateral variation of A-scan profiles to show two parameters of SS-OCT signal for delineating an oral cancer lesion. One of the parameters is the decay constant in the exponential fitting of the SS-OCT signal intensity along depth. This decay constant decreases as the A-scan point moves laterally across the margin of a lesion. The other parameter is the standard deviation of the SS-OCT signal intensity fluctuation in an A-scan. This parameter increases significantly when the A-scan point is moved across the transition region between the normal and abnormal portions. Such parameters are useful for determining the margins of oral cancer.