Review of Scientific Instruments · 2012 · 11 citations · 8 references
EngineeringElectrical EngineeringPhotonicsSpectroscopyThermal AnalysisScattering TheoryTemperature MeasurementTemperature MetrologyElectron ScatteringThomson Scattering
The large helical device Thomson scattering system was designed for the target electron temperature (T(e)) range, T(e) = 50 eV-10 keV. Above 10 keV, the experimental error becomes rapidly worse. In order to obtain reliable T(e) data in the temperature range above 10 keV, we are planning to extend the measurable T(e) range by following two methods. First we have installed one more wavelength channel that observes shorter wavelength region in polychromators. Next applying forward scattering configuration is another candidate. We estimate the data quality when the two methods are used. Both of the two methods are expected to improve T(e) data quality at T(e) ≥ 10 keV.
8
Design and performance of the Thomson scattering diagnostic on LHD
K. Narihara, I. Yamada, H. Hayashi et al. · Review of Scientific Instruments · 2001 · 178 citations · Full text
Recent Progress of the LHD Thomson Scattering System
I. Yamada, K. Narihara, H. Funaba et al. · Fusion Science & Technology · 2010 · 106 citations
Applications of phase conjugate mirror to Thomson scattering diagnostics (invited)
T. Hatae, O. Naito, M. Nakatsuka et al. · Review of Scientific Instruments · 2006 · 40 citations
Natural Sciences, Physics, Optics +15