A New Approach to Evaluate Vascular Access in HemodialysisPatients

Dammar Bouchouareb, Adriana Saveanu, Jean Michel Bartoli, M Olmer

Artificial Organs · 1998 · 26 citations · 12 references

DOIFull text

Open access

Concepts

Abstract

We evaluated the importance of vascular access in hemodialysis patients using noninvasive methods with the Transonic Systems monitor in 108 patients. Most of these patients (84%) had native vein fistulas. We found that a blood flow rate of below 500 ml/min suggested the occurrence of vascular stenosis and justified confirmation by angiography. Increased recirculation could be evaluated readily and was detected in only 10% of patients. Finally, employing the evaluation of the Kt/V index, we found a good correlation between low flux through the fistula and a low Kt/V value.

References

12