Design and performance of a reflectron based time-of-flight secondary ion mass spectrometer with electrodynamic primary ion mass separation

E. Niehuis, T. Heller, H. Feld, A. Benninghoven

Journal of Vacuum Science & Technology A Vacuum Surfaces and Films · 1987 · 137 citations · 0 references

Concepts

TL;DR

In SIMS, the number of secondary ions is limited by sample size, especially in surface reaction, trace analysis, and imaging studies, so a time‑of‑flight mass spectrometer is advantageous because it detects all masses simultaneously, offers unlimited mass range, and provides high transmission. The authors aim to develop a high‑resolution reflectron‑based TOF‑SIMS that incorporates electrodynamic mass separation and beam chopping via a pulsed 90° deflection of the primary ion beam. The instrument achieves second‑order energy focusing with a two‑stage reflectron and uses electrodynamic mass separation with a pulsed 90° deflection of the primary ion beam to chop the beam. The system attains a primary ion pulse width below 1.5 ns, a mass resolution of 13 000, and a dynamic range covering five orders of magnitude.

Abstract

In many secondary ion mass spectrometry (SIMS) investigations, the total number of generated secondary ions is limited by the amount of sample material available. This is the case in surface reaction studies as well as in organic and inorganic trace analysis or imaging SIMS. In such cases a time-of-flight mass spectrometer has some considerable advantages: quasisimultaneous detection of all masses, unlimited mass range, and very high transmission. We have developed a high-resolution reflectron based time-of-flight secondary ion mass spectrometer with a new electrodynamic mass separation and beam chopping technique based on a pulsed 90° deflection of the primary ion beam. A primary ion pulse width of less than 1.5 ns has been obtained. Second-order energy focusing in the flight path of the secondary ions is achieved by a two-stage reflectron. A mass resolution m/Δm=13 000 and a dynamic range of five orders of magnitude have been obtained with this instrument.