Brazilian Archives of Biology and Technology · 2006 · 11 citations · 4 references
MedicineRadiologyRadiation DosimetryRadiation MeasurementRadiation PhysicsDose CalibrationDosimetry InstrumentationExternal Beam DosimetryIonization Chamber DosimetryX-ray DetectorsX-ray OpticsDosimeter CalibrationParallel-plate Ionization Chamber DosimetryParallel-plate Ionization ChambersX Ray Dosimetry CalibrationX-ray DiagnosticsX-ray DosimetryX-ray Exposure Assessment
The performance of four X-ray qualities generated in a Pantak X-ray machine operating at 30-100 kV was determined with a parallel-plate ionization chamber and a Fricke dosimeter. X-ray qualities used were those recommended by Deutsch Internationale Normung DIN 6809 and dose measurements were carried out with Plexiglas® simulators. Results have shown that the Fricke dosimeter can be used not only for soft X-ray dosimetry, but also for the maintenance of low-energy measuring systems' calibration factor.
4
THE INTERNATIONAL COMMISSION ON RADIATION UNITS AND MEASUREMENTS
A. Wambersie · Journal of the ICRU · 2004 · 874 citations