Simplified capillary electrophoresis nanospray sheath‐flow interface for high efficiency and sensitive peptide analysis

Roza Wojcik, Oluwatosin O. Dada, Martin Sadı́lek, Norman J. Dovic̀hi

Rapid Communications in Mass Spectrometry · 2010 · 233 citations · 23 references

Concepts

Abstract

We report a simple nanospray sheath-flow interface for capillary electrophoresis. This interface relies on electrokinetic flow to drive both the separation and the electrospray; no mechanical pump is used for the sheath flow. This system was interfaced with an LCQ mass spectrometer. The best results were observed with a 2-microm diameter emitter tip and a 1-mm spacing between the separation capillary tip and the emitter tip. Under these conditions, mass detection limits (3sigma) of 100 amol were obtained for insulin receptor fragment 1142-1153. The separation efficiency exceeded 200,000 plates for this compound. The relative standard deviation generated during continual infusion of a 50 microM solution of angiotensin II was 2% for the total ion count and 3% for the extracted ion count over a 40-min period. Finally, the interface was also demonstrated for negative ion mode.