Micro-electrospray mass spectrometry: Ultra-high-sensitivity analysis of peptides and proteins

Mark R. Emmett, Richard M. Caprioli

Journal of the American Society for Mass Spectrometry · 1994 · 508 citations · 14 references

Concepts

TL;DR

The authors optimized a micro‑electrospray source for 300–800 nL/min flow, integrating C18 liquid‑chromatography packing into a capillary needle to enable direct, high‑sensitivity spraying. The micro‑electrospray source achieved femtomole sensitivity, detecting 1 fmol methionine enkephalin with S/N > 5:1, generating tandem MS from 500 amol, and measuring protein masses with 20–200 ppm accuracy at 2.5–200 fmol injections.

Abstract

A "micro-electrospray" ionization source has been developed that markedly increases the sensitivity of the conventional electrospray source. This was achieved by optimization of the source to accommodate nanoliter flow rates from 300 to 800-nL/min spraying directly from a capillary needle that, for the analysis of peptides, contained C18 liquid chromatography packing as an integrated concentration-desalting device. Thus, a total of 1 fmol of methionine enkephalin was desorbed from the capillary column spray needle, loaded as a 10-μL injection of 100-amol/μL solution. The mass spectrum showed the [M + H](+) ion at m/z 574.2 with a signal-to-noise ratio of better than 5:1 from a chromatographic peak with a width of about 12 s. A narrow range (15-u) tandem mass spectrum was obtained for methionine enkephalin from the injection of 500 amol, and a full-scan tandem-mass spectrum was obtained from 50 fmol. For proteins, the average mass measurement accuracy was approximately 100-200 ppm for the injection of 2.5 fmol of apomyoglobin and 20-40 ppm for 200 fmol. Carbonic anhydrase B and bovine serum albumin showed similar mass measurement accuracies.

References

14