Applied Optics · 1996 · 138 citations · 10 references
EngineeringElectrical EngineeringEquipmentInstrumentationMetrologyCalibrationEquipment CalibrationMeasurement StandardsMeasurement UncertaintyOptical InstrumentationRadiation InstrumentationCryogenic InstrumentationInstrument Measurement StandardsPrecision Measurement TechniquesRadiometer CalibrationRadiometer Design
A high-accuracy cryogenic radiometer has been developed at the National Institute of Standards and Technology to serve as a primary standard for optical power measurements. This instrument is an electrical-substitution radiometer that can be operated at cryogenic temperatures to achieve a relative standard uncertainty of 0.021% at an optical power level of 0.8 mW. The construction and operation of the high-accuracy cryogenic radiometer and the uncertainties in optical power measurements are detailed.
10
A Cryogenic Radiometer for Absolute Radiometric Measurements
John Martin, Nigel Fox, P. J. Key · Metrologia · 1985 · 276 citations
Minnesota Mining and Manufacturing Co.
Kimberly Ryan · Datamation · 1993 · 70 citations