Effect of different spray chambers on the determination of organotin compounds by high-performance liquid chromatography–inductively coupled plasma mass spectrometry

Cristina Soler Rivas, Les Ebdon, Steve J. Hill

Journal of Analytical Atomic Spectrometry · 1996 · 46 citations · 16 references

Concepts

Abstract

One of the potential problems to overcome when coupling liquid chromatography with atomic spectrometry is the low sample transport efficiency. Often this is a consequence of the interface design, and in particular the nebulizer and spray chamber configuration. The present study reports on a comparison of several spray chambers (single-pass, double-pass and cyclone types) with respect to both design and performance characteristics. In order to evaluate performance for speciation studies, different configurations have been evaluated utilizing HPLC–ICP-MS. Various organotin compounds (monobutyltin, dibutyltin, tributyltin and triphenyltin chlorides) were coinjected and the effects on resolution, sensitivity and SNR assessed. Of the spray chambers evaluated, a cyclone spray chamber (internal volume 22 ml) with cooling jacket was found to offer the best performance and gave a transport efficiency of 7.5% without loss of chromatographic resolution and sensitivity.