Densitometric Determination of Tadalafil Citrate in Tablets: Validation of the Method

Hosiana Berniati Tampubolon, Endang Sumarlik, Setiawan Dwi Saputra, Siti Cholifah, Wiwin Farina Kartinasari, Gunawan Indrayanto

Journal of Liquid Chromatography & Related Technologies · 2006 · 18 citations · 15 references

Concepts

Abstract

Abstract A simple and rapid densitometric method has been developed for determination of tadalafil citrate in tablets and its dissolution media. After extracting the samples with acetone, the solutions were spotted onto pre‐coated silica gel TLC plates which were eluted with a mixture of n‐hexane‐ethyl acetate‐methanol (8.0:6.0:2.0, v/v). quantitative evaluation was performed by measuring the absorbance reflectance of the tadalafil citrate spots at λ=285 nm. The TLC‐densitometric method is cheap, selective, precise, and accurate and can be used for routine analysis of tablets in pharmaceutical industry quality control laboratories.