A comparison of the effects of aspirin on bleeding time measured using the Simplate<sup>™</sup> method and closure time measured using the PFA‐100<sup>™</sup>, in healthy volunteers

101

Citations

0

References

1997

Year

Abstract

The change in closure time using the PFA-100 is as sensitive and reproducible to the effects of aspirin on platelet function as is the template bleeding time test. However, the PFA-100 produced less variable effects with fewer false positive results.