Electrical conductivity imaging using a variational method in <i> B <sub>z</sub> </i> -based MREIT

Oh In Kwon, Chunjae Park, Eun‐Jae Park, Jin Keun Seo, Eung Je Woo

Inverse Problems · 2005 · 19 citations · 15 references

Concepts

Abstract

Magnetic resonance electrical impedance tomography (MREIT) is a new conductivity imaging modality that was motivated to deal with the well-known ill-posedness problem in electrical impedance tomography (EIT). In order to bypass this ill-posed nature, MREIT takes advantage of an MRI scanner as a tool to capture the z-component Bz of the induced internal magnetic flux density B = (Bx, By, Bz) due to an injection current. Here, z is the direction of the main magnetic field of the MRI scanner. In this work, we propose an enhanced version of the variational gradient Bz algorithm that benefits from a careful use of boundary conditions in a variational formulation of the Bz-based MREIT model. We found that the proposed algorithm is advantageous especially in dealing with conductivity images near the boundary of the subject. Improvements in reconstructed image quality are shown by numerical simulations.