IEEE Antennas and Wireless Propagation Letters · 2012 · 26 citations · 13 references
EngineeringElectrical EngineeringHeat TransferInstrumentationMicrowave EngineeringBiomedical InstrumentationElectrical CharacterizationElectrical MeasurementElectromagnetic CompatibilityElectromagnetic MethodsMicrowave MeasurementsAbsorption Of Electromagnetic EnergyElectric Field DistributionElectric Field EffectsFinite Difference Time Domain ModelingMicrowave AbsorptionMicrowave DiagnosticsTemperature SensitivityTime Domain Electromagnetic Sensing
In this letter, the temperature measurement ability of an electrooptic probe as well as specific absorption rate (SAR) assessments via simultaneous in situ temperature and electric field characterization are reported. The measurements are carried out at 1800 MHz in a Petri dish filled with a water solution and placed in a transverse electromagnetic (TEM) cell. From the temperature sensitivity measurements, a standard deviation of 27 mK is obtained. The SAR values obtained both via temperature and electric field are also compared to finite-difference time-domain (FDTD) simulated numerical results. A difference of 5% is obtained between the two experimental SAR values. These measured SAR values are consistent with those obtained by the numerical simulations.
13
Thermal dosimetry and temperature measurements.
Douglas A. Christensen · PubMed · 1979 · 48 citations