Metrologia · 1996 · 21 citations · 7 references
EngineeringNuclear PhysicsMeasurementClock SynchronizationSignal IntegrityType B EvaluationTime DisseminationClock RecoveryUncertainty QuantificationCalibrationTiming AnalysisInstrumentationVertical Caesium BeamReliabilityPrecision MeasurementPhysicsComputer EngineeringAtomic PhysicsTime MetrologyComputer ScienceExperimental Nuclear PhysicsPrimary Atomic ClockNatural SciencesApplied PhysicsStandard Uncertainty
The CS3, a primary atomic clock with a vertical caesium beam of the Physikalisch-Technische Bundesanstalt, has been in operation since autumn 1992. A type B evaluation of its uncertainty was performed and a combined standard uncertainty of 14 × 10-15 was obtained. The clock can now, together with others, be used to realize International Atomic Time.
7
Stark Effect on Cesium-133 Hyperfine Structure
R. D. Haun, Jerrold R. Zacharias · Physical Review · 1957 · 76 citations
Stark Effect, Engineering, Physics +13
The accuracy evaluation of NIST-7
W.D. Lee, J.H. Shirley, John P. Lowe et al. · IEEE Transactions on Instrumentation and Measurement · 1995 · 48 citations