Analytical Chemistry · 2003 · 26 citations · 14 references
Natural SciencesChemistryAnalytical ChemistryAtmospheric ChemistrySpectroscopyAtmospheric Trace GasesFourier Transform SpectroscopyGas ChromatographyInfrared SpectroscopyVolatile Organic CompoundsAnalytical Detection LimitsAtmospheric Trace Gas ChemistryEnrichmentInfrared SpectraReference Method ValidationThermal Desorption AnalysisTrace Gas AnalyzersTrace Gas Detection
The determination of organic trace gases in the ambient environment at the lower ppb level is demonstrated based on a novel technique combining sorption tube sampling on Molsieve and Carbosieve S-III, thermal desorption, and detection of the trace analyte by hollow waveguide Fourier transform infrared (HWG-FT-IR) spectroscopy. While ethene concentrations of approximately 5 ppm can be directly observed using HWG-FT-IR, enrichment factors of up to 5000 were achieved by sorption tube sampling and thermal desorption. Detection limits of approximately 1 ppb are reported. Efficient enrichment by the sampling tube is achieved due to the favorable internal volume ( approximately 0.4 cm(3) at a length of 470 mm) of the hollow waveguide serving as a miniaturized gas cell. This new method was validated for ethene by thermodesorption-cryofocusing-GC-FID as the reference method. Analytical performance has been compared for standard gas mixtures and for ethene measurements in urban air. Finally, ethene data from a sampling campaign at two alpine sites in Tyrol/Austria are presented.
14
S. Hayward, C. N. Hewitt, J. H. Sartin et al. · Environmental Science & Technology · 2002 · 117 citations
Armin Hansel, Alfons Jordan, C. Warneke et al. · Plasma Sources Science and Technology · 1999 · 79 citations