IEEE Transactions on Nuclear Science · 2013 · 21 citations · 15 references
EngineeringChemistryNuclear EngineeringNuclear InstrumentationSpectroscopic AnalysisSpectroscopyGamma-ray Data AnalysisGamma-ray DetectionGamma-ray SpectroscopyNuclear SpectroscopyPrecision SpectroscopyRadiation SpectroscopyDetector Energy ResolutionHigh-purity Germanium DetectorsIsotopic AnalysisIsotopic Composition AnalysisIsotopic Ratio DeterminationMicrocalorimeter DetectorsPeak FittingSpectral Peak Deconvolution
Microcalorimeter detectors provide unprecedented energy resolution for gamma-ray spectroscopy. One application is measuring the isotopic composition of plutonium-bearing samples by non-destructive gamma-ray spectroscopy to support nuclear safeguards and nonproliferation efforts. When measured with conventional high-purity germanium (HPGe) detectors, data from these samples contain significant peak overlaps requiring spectral deconvolution for analysis. The improved energy resolution of the microcalorimeter detector reduces peak overlaps leading to improvement in the statistical error component of the total measurement uncertainty. In this paper, we describe analysis code that was developed for spectral peak fitting and isotopic content determination from microcalorimeter and HPGe data. We apply the code to data collected from several plutonium standards to quantify the improvement of the statistical error derived from the improved energy resolution.