Publication | Open Access
High Risk HPV Contamination of Endocavity Vaginal Ultrasound Probes: An Underestimated Route of Nosocomial Infection?
108
Citations
15
References
2012
Year
Our study reveals that a considerable number of ultrasound probes are contaminated with human and HR-HPV DNA, despite LLD disinfection and probe cover. In all hospitals, where LLD is performed, the endovaginal ultrasound procedure must therefore be considered a source for nosocomial HR-HPV infections. We recommend the stringent use of high-level disinfectants, such as glutaraldehyde or hydrogen peroxide solutions.