Evaluation of Ultrafiltration for the Free‐Fraction Determination of Single Photon Emission Computed Tomography (SPECT) Radiotracers: β‐CIT, IBF, and Iomazenil

Mitchell S. Gandelman, Ronald M. Baldwin, Sami S. Zoghbi, Yolanda Zea‐Ponce, Robert B. Innis

Journal of Pharmaceutical Sciences · 1994 · 195 citations · 28 references

DOIFull text

Open access

Concepts

TL;DR

An ultrafiltration system was evaluated at 25 °C to determine the free fraction of β‑CIT, IBF, and iomazenil for functional brain SPECT imaging. The technique yielded <9 % relative error across temperature, storage, centrifugal force, tracer concentration, and filtration percentage, <5 % deviation from equilibrium dialysis for β‑CIT and iomazenil (16 % for IBF), <6 % precision, and faster, easier operation with strong correlation to dialysis.

Abstract

An ultrafiltration system was evaluated for the free‐fraction measurement of SPECT radiotracers (β‐CIT, IBF, and iomazenil) used in functional brain imaging. The effect of temperature, storage, centrifugal force, tracer concentration, and percentage filtered demonstrated a relative error of < 9%. As a result of the minimal temperature effect, 25 °C was employed for all measurements. A comparison of the ultrafiltration system with equilibrium dialysis revealed < 5% difference for β‐CIT and iomazenil, but 16% for IBF. Additionally, the time and ease of operation considerably favored the ultrafiltration system. The precision quantitated by repetition was < 6% for between‐run and within‐run variability. In conclusion, ultrafiltration provided rapid results, demonstrated minor analytical errors, revealed generally good correlation with equilibrium dialysis, and allowed excellent precision.