Analytical Chemistry · 1999 · 142 citations · 39 references
Natural SciencesBiochemistryChemistryAnalytical ChemistryMass Spectrometric AnalysisSpectroscopyIonization TechniquesMass AnalyzersMass SpectrometryCharge State DistributionsElectrospray IonizationMass Spectrometry CharacterizationMatrix-assisted Laser Desorption Ionization Time-of-flight Mass SpectrometryMatrix-assisted Laser Desorption/ionizationMatrix-assisted Laser Desorption/ionization Time-of-flight Mass SpectrometryNeutralizationProton Transfer Reactions
Charge reduction electrospray mass spectrometry (CREMS) is introduced as a new technique for analyzing complex biological mixtures by producing predominantly singly charged ions. CREMS achieves charge reduction by reacting electrospray‑generated multiply charged ions with bipolar ions produced from a 210Po α‑particle source in a neutralization chamber before the mass spectrometer, and the resulting ions are detected on an orthogonal electrospray time‑of‑flight instrument, with the chamber adaptable to any analyzer. The method shows a size‑dependent signal drop‑off for oligonucleotides comparable to MALDI, and proton‑transfer reactions are identified as the mechanism reducing charge on proteins and oligonucleotides in both positive and negative ion modes.
A new mass spectrometric technique, charge reduction electrospray mass spectrometry (CREMS), allowing the analysis of complex mixtures of biological molecules is described. The charge state of ions produced by electrospray ionization may be reduced in a controlled manner to yield predominantly singly charged ions through reactions with bipolar (i.e., both positively and negatively charged) ions generated using a 210Po α particle source. The electrospray-generated multiply charged ions undergo charge reduction in a "neutralization chamber" positioned before the entrance nozzle to the mass spectrometer. The ions are detected using a commercial orthogonal electrospray time-of-flight mass spectrometer, although the neutralization chamber can be adapted to virtually any mass analyzer. The CREMS results obtained exhibit a signal intensity drop-off with increasing oligonucleotide size similar to that observed with matrix-assisted laser desorption/ionization mass spectrometry. Proton-transfer reactions were found to be responsible for reducing charge on proteins and oligonucleotides in both positive and negative ion mode.
39