Pressure Measurement Made by the Utilization of Ruby Sharp-Line Luminescence
772
Citations
7
References
1972
Year
Mineral PhysicEngineeringMeasurementMicroscopyOptical TestingLuminescence PropertyReciprocal CentimetersOptical PropertiesOptical DiagnosticsInstrumentationMaterials SciencePhysicsPrecision Pressure MeasurementRuby SampleDiamond-like CarbonNatural SciencesSpectroscopyApplied PhysicsRuby Sharp-line Luminescence
A rapid, convenient technique for precision pressure measurement in the diamond-anvil high-pressure cell, which makes use of the sharp-line (R-line) luminescence of ruby, has been developed. The observed shift is -0.77 +/-0.03 reciprocal centimeters per kilobar for R(1) and -0.84+/- 0.03 reciprocal centimeters per kilobar for R(2) to lower energy and is approximately linear in the range studied (to 22 kilobars). Line-broadening has been observed in some instances and has been tentatively identified with nonhydrostatic conditions surrounding the ruby sample.