Analytical Letters · 2012 · 45 citations · 23 references
Natural SciencesChemistryAnalytical ChemistryMass Spectrometric AnalysisSpectroscopyDetector SystemsInstrumental AnalysisIonization TechniquesMass AnalyzersMass SpectrometryExplosive Detection SystemsExplosivesExplosives AnalysisExplosives ChemistryExplosives DetectionIonizationMass Spectrometric DetectionPortable Analytical Devices
A small low-temperature plasma (LTP) ionization probe was coupled to a portable mass spectrometer for the rapid detection of trace explosives on surfaces. Using only a small diaphragm pump to supply ambient air to the LTP source, 100 ng each of pentaerythritol tetranitrate (PETN), 1,3,5-trinitroperhydro-1,3,5-triazine (RDX), and 2,4,6-trinitrophenylmethylnitramine (Tetryl) were detectable on glass in under 1 minute. The main ion signal from these molecules (M) is the [M + NO3]− species. While much optimization remains, it is believed that this miniature LTP source will remove the need for external gas cylinders and additional heating for in situ explosives detection using portable mass spectrometers.
23
Low-Temperature Plasma Probe for Ambient Desorption Ionization
Jason D. Harper, Nicholas A. Charipar, Christopher C. Mulligan et al. · Analytical Chemistry · 2008 · 732 citations
Natural Sciences, Chemistry, Physics +17
Ismael Cotte-Rodríguez, Zoltán Takáts, Nari Talaty et al. · Analytical Chemistry · 2005 · 345 citations
Field detection and monitoring of explosives
Jehuda Yinon · TrAC Trends in Analytical Chemistry · 2002 · 284 citations · Full text