Instrumentation Science & Technology · 2006 · 21 citations · 27 references
Natural SciencesChemistryPharmaceutical AnalysisAnalytical ChemistryMass Spectrometric AnalysisSpectroscopyBioanalytical ChemistryBioanalytical Method ValidationDetection Limit AssessmentInductively Coupled Plasma Optical Emission SpectrometryInductively Coupled Plasma Optical Emission SpectroscopyInductively Coupled Plasma SpectrometryInstrument Detection LimitMethod Detection LimitOptical Emission SpectroscopyUltrasonic Nebulization
Abstract An on‐line gadolinium preconcentration and determination system, implemented with ultrasonic nebulization coupled to inductively coupled plasma optical emission spectrometry (ICP‐OES), associated with flow injection (FI) was studied. Gadolinium was retained as Gd‐2‐(5‐bromo‐2‐pyridilazo)‐5‐diethylaminophenol complex (Gd‐5‐Br‐PADAP) at pH 9.0, on the inner walls of the knotted reactor (KR). The collected analyte complexes were quantitatively eluted from the 200 cm KR with 30% (v/v) nitric acid. An enhancement factor of 255‐fold was obtained (17 for KR and 15 for USN). The detection limit (DL) value for the preconcentration of 15 mL of sample was 4.0 ng L−1. The relative standard deviation (RSD) was 3.5%, calculated from the peak heights obtained. The calibration graph using the preconcentration system for gadolinium was linear, with a correlation coefficient of 0.9996 at levels near the detection limit up to at least 200 µg L−1. The method was successfully applied to the determination of gadolinium in urine samples.
27
Elissaveta Ivanova, Karima Benkhedda, F. Adams · Journal of Analytical Atomic Spectrometry · 1998 · 88 citations
Steffen Nielsen, Elo Harald Hansen · Analytica Chimica Acta · 1998 · 76 citations