Review of Scientific Instruments · 2007 · 58 citations · 36 references
EngineeringBioengineeringBiomedical EngineeringBiosensorsSpectroscopic AnalysisSpectroscopyBiomedical InstrumentationBiopotential MeasurementElectrical MeasurementImpedance SpectroscopyMicrofluidic BiosensorsMicrofluidicsBandwidthBroadband Dielectric SpectroscopyBroadband SpectroscopyDielectric SpectroscopyFrequency-dependent Electrical MeasurementsImpedance AnalysisImpedance MeasurementMicrofluidic Channel Networks
A maximum length sequence (MLS) is used to perform broadband impedance spectroscopy on a dielectric sample. The method has a number of advantages over other pulse-based or frequency sweep techniques. It requires the application of a very short sequence of voltage steps in the microsecond range and therefore allows the measurement of time-dependent impedance of a sample with high temporal resolution over a large bandwidth. The technique is demonstrated using a time-invariant passive RC network. The impedance of single biological cell flowing in a microfluidic channel is also measured, showing that MLS is an ideal method for high speed impedance analysis.
36
<i>Physical Techniques in Biological Research</i>
William L. Nastuk, Joseph G. Hoffman · Physics Today · 1963 · 937 citations
Pseudo-random sequences and arrays
F. MacWilliams, N. J. A. Sloane · Proceedings of the IEEE · 1976 · 737 citations