Review of Scientific Instruments · 1955 · 13 citations · 2 references
Natural SciencesPhysicsElectromagneticsElectromagnetismOpticsChromatic AberrationElectron MicroscopyOptical AberrationsTransmission Electron MicroscopyChromatic Aberration CorrectionElectron Beam OpticsMicroscope Chromatic Correction
Experimental results on the reduction of chromatic aberration in the magnetic electron microscope are described. It is proved that the chromatic field aberration can be completely compensated by a proper combination of two lenses with adequately designed pole pieces and the voltage stabilization tolerance is greatly improved. Micrographs with a resolution of about 5 mμ can be obtained even under a voltage fluctuation as large as ΔE/E=1.10−2.
2