Journal of Microscopy · 1998 · 31 citations · 7 references
Natural SciencesBiophysicsBiomedical ImagingBiophotonicsOpticsFluorescence ImagingFluorescence MicroscopyImage QualityImage Quality AssessmentLight MicroscopyOptical ImagingFluorescence Microscopy QuantificationIntrinsic NoiseLight ThroughputPhotobleaching AnalysisSignal-to-noise RatioSignal-to-noise Ratio AnalysisThroughputTime-resolved Fluorescence Imaging
A simple means of evaluating the performance of fluorescence microscopes is described. The proposed test gives an overall figure of merit that takes into account all of the important instrumental parameters that determine image quality. The essence of the test is to use a specimen whose photobleaching rate is a measure of the illumination time-intensity integral. When this time-intensity integral is controlled, the signal-to-noise ratio in the image of small subresolution objects becomes a system-independent measure of light throughput, resolution and intrinsic noise.
7
David R. Sandison, David W. Piston, Rebecca M. Williams et al. · Applied Optics · 1995 · 89 citations
Natural Sciences, Biomedical Engineering, Biomedical Imaging +15
Chapter 13 Basic Fluorescence Microscopy
D. Lansing Taylor · Methods in cell biology · 1988 · 79 citations