Measurement Science and Technology · 2001 · 32 citations · 18 references
Natural SciencesPhysicsOpticsPhotonicsRadiation DetectionUltrafast OpticsX-ray Data AnalysisX-ray DetectionX-ray DetectorsX-ray OpticsX-ray TimingDynamic RangeTime ResolutionTime-resolved Detection
An x-ray streak camera system uses a laser triggered photo-conductive switch to synchronize the high voltage ramp applied to its sweep plates to the photo-excitation of a sample. This technique allows the jitter between successive sweeps to be kept below 100 fs compared to over 1 ps using classical methods. By accumulating a stable pulsed x-ray signal over many sweeps, a very high dynamic range can be achieved whilst maintaining a sub-picosecond time resolution. The ultimate time resolution is limited by the dispersion of electrons in the streak tube. Whilst triggering with a laser at 900 Hz we have achieved a time resolution (FWHM) of 640 fs at a detection wavelength of 267 nm for a 60 s accumulation period, corresponding to over 50 000 sweeps.
18
Picosecond–milliångström lattice dynamics measured by ultrafast X-ray diffraction
Christoph Rose-Petruck, Ralph Jimenez, Ting Guo et al. · Nature · 1999 · 516 citations
Ultrafast x-ray absorption probing of a chemical reaction
F. Ráksi, Kent R. Wilson, Zhiming Jiang et al. · The Journal of Chemical Physics · 1996 · 184 citations
Natural Sciences, Chemistry, Physics +17
High-power switching with picosecond precision
G. Mourou, Wayne H. Knox · Applied Physics Letters · 1979 · 124 citations