Physiological Measurement · 2008 · 61 citations · 8 references
EngineeringBioengineeringBiomedical EngineeringBiomedical Signal ProcessingBiosensorsSpectroscopyBiomedical InstrumentationImpedance SpectroscopyBroadband MeasurementElectrical PropertiesImpedance MeasurementImpedance MonitoringSignal-to-noise RatioSignal-to-noise Ratio Analysis
Frequency domain impedance measurements are still the common approach in assessing passive electrical properties of cells and tissues. However, due to the time requirements for sweeping over a frequency range for performing spectroscopy, they are not suited for recovering fast impedance changes of biological objects. The use of broad bandwidth excitation and monitoring the response as a function of time will greatly reduce the measurement time. The widespread usage of a square wave excitation is simple but not always the best choice. Here we consider different waveforms for excitation and discuss not only the advantages but also their limitations. Measurements in a miniaturized chamber where frequency and time domain measurements are compared show the suitability of different waveforms as excitation signals for the measurements of bio-impedance. The chirp excitation has been found to be most promising in terms of frequency range, signal-to-noise ratio and crest factor.