MedicineRadiologyBiomedical ImagingDiagnostic ImagingDiagnostic RadiologyHead ImagingMagnetic Resonance Image ReconstructionMagnetic Resonance ImagingMagnetic Resonance Imaging PhysicsMedical ImagingMotion CorrectionFast Spin EchoMotion ArtifactsPhase CorrelationRigid Motion CorrectionSpin Echo ImagingTranslational MotionTurbo Spin Echo
The use of phase correlation to detect rigid-body translational motion is reviewed and applied to individual echotrains in turbo-spin-echo data acquisition. It is shown that when the same echotrain is acquired twice, the subsampled correlation provides an array of delta-functions, from which the motion that occurred between the acquisitions of the two echotrains can be measured. It is shown further that a similar correlation can be found between two sets of equally spaced measurements that are adjacent in k-space. By measuring the motion between all adjacent pairs of k-space subgroups, the complete motion history of a subject can be determined and the motion artifacts in the image can be corrected. Some of the limiting factors in using this technique are investigated with turbo-spin-echo head and hand images.
28
SENSE: Sensitivity encoding for fast MRI
Klaas P. Pruessmann, Markus Weiger, Markus B. Scheidegger et al. · Magnetic Resonance in Medicine · 1999 · 6.1K citations