Magnetic Resonance in Medicine · 2003 · 40 citations · 22 references
Natural SciencesChemistryInstrumentationSequence AnalysisSpectroscopyMeasurement System AnalysisSample PreparationError ReductionInversion Recovery SequencesLook-locker SequencesOptimization-based InversionRandom ErrorsSystematic ErrorsT1 Mapping SequencesVariable Flip Angle Sequences
A methodology is presented for the reduction of both systematic and random errors in T(1) determination using TAPIR, a Look-Locker-based fast T(1) mapping technique. The relations between various sequence parameters were carefully investigated in order to develop recipes for choosing optimal sequence parameters. Theoretical predictions for the optimal flip angle were verified experimentally. Inversion pulse imperfections were identified as the main source of systematic errors in T(1) determination with TAPIR. An effective remedy is demonstrated which includes extension of the measurement protocol to include a special sequence for mapping the inversion efficiency itself.
22
Fast and precise T1 imaging using a TOMROP sequence
Gunnar Brix, Lothar R. Schad, M. Deimling et al. · Magnetic Resonance Imaging · 1990 · 178 citations
A fast 3D Look-Locker method for volumetric T1 mapping
Elizabeth Henderson, Graeme C. McKinnon, Ting-Yim Lee et al. · Magnetic Resonance Imaging · 1999 · 142 citations