Measurement of Electron Density and Effective Atomic Number by Dual-Energy Scan Using a 320-Detector Computed Tomography Scanner with Raw Data-Based Analysis

35

Citations

10

References

2014

Year

Abstract

We evaluated the accuracy of the electron densities and effective atomic numbers determined by raw data-based dual-energy analysis on a 320-detector computed tomography scanner. The mean (SD) errors between the measured and true electron densities and between the measured and true effective atomic numbers were 1.3% (1.5%) and 3.1% (3.2%), respectively. Electron densities and effective atomic numbers can be determined with high accuracy, which may help to improve accuracy in radiotherapy treatment planning.