Measurement of Electron Density and Effective Atomic Number by Dual-Energy Scan Using a 320-Detector Computed Tomography Scanner with Raw Data-Based Analysis
35
Citations
10
References
2014
Year
EngineeringEffective Atomic NumberPositron Emission TomographyElectron MicroscopyElectron SpectroscopyRadiotherapy Treatment PlanningRadiation Therapy PlanningPhoton-counting Computed TomographyRadiation OncologyNuclear MedicineRadiologyHealth SciencesElectron DensityMedical ImagingPhysicsProton TherapyDosimetryHigh AccuracySpectroscopyApplied PhysicsElectron MicroscopeDual-energy ScanElectron Densities
We evaluated the accuracy of the electron densities and effective atomic numbers determined by raw data-based dual-energy analysis on a 320-detector computed tomography scanner. The mean (SD) errors between the measured and true electron densities and between the measured and true effective atomic numbers were 1.3% (1.5%) and 3.1% (3.2%), respectively. Electron densities and effective atomic numbers can be determined with high accuracy, which may help to improve accuracy in radiotherapy treatment planning.