Ultrahigh-resolution full-field optical coherence tomography

Arnaud Dubois, Kate Grieve, Gaël Moneron, Romain Lecaque, Laurent Vabre, Claude Boccara

Applied Optics · 2004 · 382 citations · 61 references

Concepts

TL;DR

The authors developed a white‑light interference microscope for ultrahigh‑resolution full‑field optical coherence tomography of biological media. The system uses a Linnik‑type interferometer illuminated by a tungsten halogen lamp, with en face images reconstructed from interferometric data captured by a high‑speed CCD camera. The microscope achieves 1.8 µm × 0.9 µm transverse‑axial resolution, 90 dB shot‑noise‑limited sensitivity in 4 s per image, enabling subcellular imaging of plant, animal, and human tissues.

Abstract

We have developed a white-light interference microscope for ultrahigh-resolution full-field optical coherence tomography of biological media. The experimental setup is based on a Linnik-type interferometer illuminated by a tungsten halogen lamp. En face tomographic images are calculated by a combination of interferometric images recorded by a high-speed CCD camera. Spatial resolution of 1.8 μm × 0.9 μm (transverse × axial) is achieved owing to the extremely short coherence length of the source, the compensation of dispersion mismatch in the interferometer arms, and the use of relatively high-numerical-aperture microscope objectives. A shot-noise-limited detection sensitivity of 90 dB is obtained in an acquisition time per image of 4 s. Subcellular-level images of plant, animal, and human tissues are presented.