The Analyst · 2008 · 14 citations · 14 references
Natural SciencesChemistryMaterials ScienceOptical MaterialsSensorsSpectroscopyChemical SensorsFiber OpticsLigand BindingOptical SensingOptical SensorsOptical SpectroscopyChemical Optical SensorsChemical SensingChemical Sensing SystemsOptical Chemical SensingOptical DetectionOptical Extinction CoefficientOptical Fiber SensorsSpectral Interferences
Detection of radiological materials in the solution phase is restricted by conventional radiation-counting techniques owing to extreme attenuation. Chemical sensing of the resultant radiological species such as uranyl UO(2)(2+) is possible on the surface of a plastic or glass fibre optic. A dihydroxy isoamethryin complex is tethered to the fibre surface which has a large extinction coefficient (119 000 M(-1) cm(-1) at lambda = 439 nm) and changes colour upon binding UO(2)(2+). The spectral changes are greater on the surface than in solution and binding is specific to UO(2)(2+) with small interferences from Gd(3+). Monitoring the spectral response in three detector bands in the red, green and blue enable the optical power change to be measured with sensitivities of 1 mdB, allowing UO(2)(2+) to be detected confidently at 50-100 ppb levels. Real-time kinetic analysis enables discrimination between the target species and possible interferents.
14
Fiber-Optic Biosensors—Trends and Advances
Mehrab Mehrvar, Chris Bis, Jeno M. Scharer et al. · Analytical Sciences · 2000 · 138 citations
Azo oxide gels for optical storage
Fréderic Chaput, D. Riehl, Yves Lévy et al. · Chemistry of Materials · 1993 · 81 citations