Magnetic Resonance in Medicine · 2005 · 31 citations · 19 references
MedicineNeurobiologyNeuroscienceCognitive NeuroscienceMagnetic Resonance BiophysicsNeurochemistryBrain ImagingCerebral PhysiologyMagnetic Resonance ImagingMagnetic Resonance SpectroscopyMetabolite QuantificationGlutamate NeurotransmissionNuclear Magnetic Resonance QuantificationNuclear Magnetic Resonance Spectroscopy Data Analysis
The goal of this study was to measure the interindividual reproducibility of glutamate quantification in 1.5-T (1)H MRS of human brains. To determine the effective echo time (TE) for glutamate quantification, spectra from a phantom and 12 participants were obtained with TE = 30, 35, 40, and 144 ms (repetition time (TR) = 2000 ms and volume of interest = 4 cm(3)). The average Cramer-Rao lower bounds for glutamate quantification using LCModel was lowest in two experiments when TE = 40 ms.Twenty-one subjects participated in experiments that measured interindividual reproducibility of glutamate quantification. Spectra were acquired with TR = 6000 ms and TE = 40 ms. Results showed that the coefficients of variance were 11.0 and 13.1% in the anterior cingulate cortex and insula, respectively. This suggests that glutamate can be reproducibly measured from 1.5-T (1)H MRS with long TR, effective TE, and the LCModel.
19