Hard X-ray Imaging Microscopy using X-ray Guide Tube as Beam Condenser for Field Illumination

Yoshio Suzuki, Akihisa Takeuchi, Kentaro Uesugi, Yasuko Terada, Hiromoto Nakazawa, Sumito Ohzawa, Tomoki Aoyama, Hajime Nii, Katsumi Handa

Journal of Physics Conference Series · 2013 · 10 citations · 5 references

DOIFull text

Open access

Concepts

Abstract

An optical system for illumination of object in x-ray imaging microscopy is developed. The optical system is a beam condenser consisting of a single-bounce conical-shape mono-capillary (x-ray guide tube: XGT) made of Pyrex glass. The XGT condenser was tested at the beam line 47XU of SPring-8 using a Fresnel zone plate as an objective lens. Comparing with the microscope without beam condenser, the flux density is improved by a factor of 12–20 in the x-ray energy range of 6–8 keV. Test patterns with a 50 nm-structure are clearly resolved at 8 keV with an exposure time less than 1 s.